• Medical Diagnostic Equipment
  • Surgical and Therapeutic Equipment
  • Medical Furniture
  • Laboratory Equipment
  • Cleaning and Sterilization

2998 Photodiode Array Detector

The Waters 2998 photodiode array detector is a high-performance optical detection instrument designed for demanding chromatography laboratories. Offering high spectral accuracy over a wide wavelength range (190 to 800 nm), it enables the detection of impurities and precise quantification of compounds. Thanks to its design and optimised software control, the 2998 is ideally suited to both routine applications and the development of complex analytical methods.



Description

The 2998 Photodiode Array Detector for precision chromatographic analysis

The Waters 2998 Photodiode Array (PDA) detector is an essential analytical solution for laboratories seeking to combine reliability, spectral performance and flexibility in their HPLC analyses. Thanks to its design and the integration of high-precision optical components, this instrument covers a wavelength range from 190 to 800 nm with a digital resolution of 1.2 nm per pixel, supported by an array of 512 photodiodes. This architecture ensures fast and complete spectral acquisition, ideal for identifying and quantifying impurities, monitoring chromatographic peaks and analysing complex spectral profiles in real time.

At the heart of the 2998 PDA is a deuterium arc lamp, which guarantees a stable and high-performance light source with a specified lifetime of 2,000 hours or one year. The exclusive TaperSlit™ technology of the flow cell optimises detection sensitivity and linearity while minimising dispersion effects, making this detector particularly effective for demanding analyses. The standard analytical cell has an optical path length of 10 mm and a volume of 8.4 µL, suitable for most standard HPLC applications.

The 2998 detector also features remarkable base stability, with noise below 10 µAU peak-to-peak and drift below 1 mAU/h under controlled conditions, ensuring reproducible analytical results. Its sampling rate of up to 80 points per second enables fast and accurate data acquisition, even under high-speed chromatography conditions. At the same time, its ability to operate in 2D and 3D modes simultaneously offers exceptional flexibility in the analysis and interpretation of results, particularly when using Waters Empower™ software.

Designed to integrate seamlessly with Alliance HPLC, ACQUITY Arc and ACQUITY Arc Bio systems, the 2998 detector offers several configurations for different analytical applications, including biocompatible chromatography and purification.

Easy to use and maintain, the 2998 detector offers a user-friendly interface with multiple analogue and event inputs and outputs for seamless integration into any laboratory environment. Its compact size, low noise level (<58 dBA) and environmental robustness (operating between 4 and 40°C) make it ideal for modern, regulated working environments.


Features

  • High-resolution PDA detector with 512 photodiodes
  • Full spectrum from 190 to 800 nm with high precision
  • Fast acquisition up to 80 Hz for high-throughput analysis
  • Simultaneous 2D and 3D analysis mode
  • Patented TaperSlit™ flow cell for improved sensitivity
  • Control via Empower™ software
  • Compatible with HPLC Alliance, ACQUITY Arc and Arc Bio
  • Erbium filter technology for precision calibration
  • Quiet and compact operation
  • User-friendly interface with multiple integration options

Technical Details

  • Wavelength range: 190 to 800 nm
  • Wavelength accuracy: ±1 nm (with patented erbium filter)
  • Wavelength repeatability: ±0.1 nm
  • Bandwidth: 1.2 nm
  • Number of photodiodes: 512
  • Digital resolution: 1.2 nm/pixel
  • Linearity: ≤5% at 2.0 AU (at 257 nm)
  • Background noise: ≤10 x 10⁻⁶ AU peak-to-peak (254 nm)
  • Drift: ≤1.0 x 10⁻³ AU/h (254 nm)
  • Sampling rate: Up to 80 points/s
  • Light source: Deuterium lamp
  • Lamp life: 2000 hours or 1 year
  • Flow cell design: Patented TaperSlit™
  • Optical path length: 10 mm (analytical cell)
  • Cell volume: 8.4 µL
  • Pressure limit: 1000 psi
  • Materials in contact: 316 stainless steel, fused silica, PEEK, fluoropolymers, MP35N, titanium
  • Dimensions: 34.3 x 20.8 x 61.0 cm (W x H x D)
  • Weight: 14.5 kg
  • Operating temperature: 4 to 40 °C
  • Humidity: 20% to 80% (non-condensing)
  • Power consumption: 195 VA
  • Power supply: 100 to 240 VAC, 50/60 Hz
  • Inputs/outputs: 4 event inputs, 2 analogue outputs, 2 event outputs
  • Noise level: <58 dBA


Compatible Accessories

  • Optional flow cells:
    • Standard analytical (8.4 µL, 10 mm) – Ref. 205001023
    • Low dispersion analytical (8.4 µL, 10 mm) – Ref. 205001552
    • Biocompatible low dispersion (8.4 µL, 10 mm) – Ref. 205001043
    • Microbore (2.7 µL, 8 mm) – Ref. 205001024
    • Semi-preparative (16.3 µL, 3 mm) – Ref. 205001025
    • Self-purifying (12.3 µL, 0.5 mm, dual channel) – Ref. 205001026

Discover more in this category

Logistics

Remma and its partners provide you with serviced and configured equipment tailored to your needs. Receive your product at your facility with standard or express delivery, and an installation option, all through a single point of contact.

Warranty

Our 12/24-month warranty ensures the safety and reliability of your medical equipment throughout its lifespan.

Flexible Financing

Settle your medical equipment in up to 36 installments with our financing service or pay after delivery with Remma Facility.