• Medical Diagnostic Equipment
  • Surgical and Therapeutic Equipment
  • Medical Furniture
  • Laboratory Equipment
  • Cleaning and Sterilization
  • Spare Parts

G1376A Capillary Pump - Good Condition

Agilent's G1376A , also known as the 1260 Infinity capillary pump, is designed for ultra-low flow liquid chromatography applications. With its electronic flow control system and two independent pump heads, it offers exceptional precision in solvent delivery. Compatible with regulated environments, it incorporates diagnostic, compressibility compensation and wear monitoring functions. This module is ideal for laboratories that demand flow stability, low dead volume and reproducible performance, particularly in capillary and nanometric analyses.



Offer Details

Accessories :

32752

Options :

Delivery time between 10-14 working days

The Offer Details are provided by the seller and automatically translated.

Description

Agilent G1376A: high-precision liquid chromatography capillary pump

The Agilent G1376A capillary pump is designed to meet the requirements of laboratories specialising in ultra-low flow liquid chromatography. It is based on a two-channel independent architecture, each equipped with two pistons in series, allowing for precise and stable solvent delivery. The electronic flow control system, combined with an electromagnetic proportioning valve (EMPV), ensures dynamic regulation of the flow to the column, while eliminating variations related to pressure or the nature of the solvents.

The internal configuration of the pump allows rapid gradients to be generated across the entire composition range, with reduced dead volume thanks to optimisation of the hydraulic path. The mobile phases are mixed in a static tube filled with stainless steel balls, ensuring constant homogeneity. The flow sensor, based on mass thermal sensitivity, provides reliable measurement calibrated according to the solvents used.

Modular design and simplified maintenance

The G1376A incorporates Agilent's E-PAC concept, combining layers of EPP foam for shock absorption and internal electromagnetic shielding. This design facilitates access to components for maintenance while reducing radio interference. The module also features an Early Maintenance Feedback (EMF) system, which tracks seal wear and solvent volume pumped, with customisable thresholds for planning maintenance.

The pump is compatible with Agilent control software (ChemStation, EZ-Chrom, OL) and offers several communication interfaces: CAN, RS-232C, LAN and APG Remote. It can be integrated into single or dual stack configurations, with optimised connectivity for other modules in the 1200/1260 Infinity series.

Performance and compliance for regulated environments

The G1376A is designed to operate in environments subject to strict standards such as GMP, ISO and FDA. It incorporates diagnostic, leak detection, compressibility compensation and automatic calibration functions. Its operating pressure range is 20 to 400 bar, with a flow accuracy of less than 0.7% RSD. The module is equipped with stainless steel filters, fused silica capillaries and components resistant to aggressive solvents, ensuring optimal durability.

With its robust design, dispensing accuracy and tracking capabilities, the G1376A is a strategic choice for laboratories seeking to control their capillary flows while complying with regulatory requirements.

Features

  • Two independent pumping channels with series pistons
  • Solvent selection valve with two inlets per channel
  • Electromagnetic proportioning valve (EMPV) with flow sensor
  • Automatic compensation for pressure and viscosity variations
  • Integrated diagnostic system with status indicators and error messages
  • Manual and automatic purge function
  • User interface via Instant Pilot or Agilent software
  • Monitoring of seal wear and pumped volume (EMF)
  • Compatible with capillary and nanometric configurations
  • Single or double stack mounting depending on laboratory layout

Technical Details

  • Column flow range: 0.01 to 20 µL/min (up to 100 µL/min with extended kit)
  • Pressure range: 20 to 400 bar
  • Flow accuracy: < 0.7% RSD or 0.03% SD
  • Retention volume: typically 3 µL (up to 12 µL depending on configuration)
  • Compressibility compensation: configurable according to solvent
  • Materials in contact with solvents: stainless steel, gold, sapphire, ruby, PEEK
  • Dimensions: 180 x 345 x 435 mm
  • Weight: 17 kg
  • Power consumption: 180 VA / 75 W
  • Operating temperature: 4 to 55 °C

Compatible Accessories

  • Module-to-module CAN cable (1 m)
  • Manual purge valve with bracket
  • Stainless steel solvent inlet filters (x4)
  • Flow sensor capillary (20 µL, 50 µm ID, 55 cm)
  • Spanner wrenches (4 mm, 1/2", 7/16")
  • Torque adapter for mounting
  • Flexible tubing (2 x 1 m)
  • Replacement frits (0.5 µm)
  • Insertion tool for capillaries

Logistics

Remma and its partners provide you with serviced and configured equipment tailored to your needs. Receive your product at your facility with standard or express delivery, and an installation option, all through a single point of contact.

Warranty

Our 12 month warranty ensures the safety and reliability of your medical equipment throughout its lifespan.

Flexible Financing

Settle your medical equipment in up to 36 installments with our financing service or pay after delivery with Remma Facility.