• Medical Diagnostic Equipment
  • Surgical and Therapeutic Equipment
  • Medical Furniture
  • Laboratory Equipment
  • Cleaning and Sterilization
  • Spare Parts

HCP50 Humidity Chamber

The Memmert HCP50 humidity-controlled climate chamber provides a stable and precise environment for accelerated life testing and 85/85 testing. With its active humidity control and fine temperature regulation, it ensures reproducible conditions for laboratories and industries. Equipped with a multifunctional heating system on six sides, an insulated stainless steel door and a TwinDISPLAY control panel, it combines reliability and performance. Its 56-litre volume, Ethernet/USB connectivity and AtmoCONTROL software make it an efficient tool for applications requiring precision and traceability.



Description

Memmert HCP50 – Humidity-controlled climate chamber for precise and reproducible testing

The Memmert HCP50 is a humidity chamber specially designed to provide a perfectly controlled environment for stability tests, life cycle tests and quality controls. It features active humidity control, allowing the range to be set from 20 to 95% RH with an accuracy of 0.5% RH. Its temperature control covers a range from +7 °C above ambient temperature to +90 °C, with an accuracy of 0.1 °C. Two Pt100 DIN Class A sensors in a 4-wire configuration provide mutual monitoring and ensure continuous operation in the event of a sensor failure.

This environmental accuracy allows real or extreme conditions to be faithfully reproduced, which is essential for 85/85 tests (85% RH at 85°C) or other standardised protocols. The SetpointWAIT system, integrated into the controller, ensures that test timing only starts once the target parameters have been reached.

Optimised design for reliability and efficiency

The HCP50 features a multifunctional heating system distributed across four sides, complemented by door and rear heating to prevent condensation. This design ensures uniform temperature and humidity distribution inside the chamber. Its 1.4301 (ASTM 304) stainless steel interior and insulated compression-locking door guarantee robustness and tightness.

Heated interior glazing limits condensation while providing visibility of the sample during testing. The chamber can accommodate up to 5 shelves, with a maximum capacity of 75 kg, allowing it to be adapted to the needs of each application. The 40 mm diameter cable entry at the rear provides a practical solution for connecting sensors or other external devices.

Control, connectivity and safety

The TwinDISPLAY ControlCOCKPIT control panel, equipped with two high-definition colour TFT screens, allows intuitive adjustment of temperature, humidity, programme duration and various time parameters. The AtmoCONTROL software facilitates the creation, transfer and management of programmes via USB or Ethernet.

Safety is enhanced by the AutoSAFETY system, an over- and under-temperature monitoring device that automatically follows the setpoint with a predefined tolerance. If the setpoint is exceeded, the heating is interrupted and a visual and audible alarm is triggered. The integrated self-diagnostic system ensures constant monitoring of the control functions.

The Memmert HCP50 meets the requirements of modern laboratories, combining precision, data traceability and long-term reliability. Its compact and robust design makes it an ideal choice for demanding test environments.

Features

  • Active humidity control for stable conditions
  • Heating distributed over 6 sides for thermal uniformity
  • Insulated stainless steel door with 2-point locking
  • Heated interior glazing for condensation-free visibility
  • SetpointWAIT function for precise cycle start-up
  • Integrated self-diagnostics for continuous monitoring
  • Visual and audible alarms in case of deviation
  • Intuitive programming via TwinDISPLAY and AtmoCONTROL
  • Convenient connections via 40 mm passage port
  • Compatible with 85/85 tests and similar protocols

Technical Details

  • Working temperature range: +7 °C above ambient to +90 °C
  • Temperature control range: +18 to +90 °C
  • Temperature control accuracy: 0.1 °C
  • Probes: 2 Pt100 Class A, 4-wire mounting
  • Humidity range: 20 to 95% RH (or RH-Off)
  • Humidity control accuracy: 0.5% RH
  • Controller: TwinDISPLAY ControlCOCKPIT with dual colour TFT display
  • Timer: 1 min to 99 days
  • Calibration: 3 free values for temperature and humidity
  • Interfaces: Ethernet LAN, USB
  • Software: AtmoCONTROL
  • Heating system: 6 sides with door and rear heating
  • Useful volume: 56 L
  • Internal dimensions: 400 × 425 × 330 mm
  • Max. load capacity: 75 kg (15 kg per shelf)
  • External dimensions: 559 × 795 × 521 mm
  • Net weight: 55 kg
  • Power supply: 230 V or 115 V, 50/60 Hz, approx. 1520 W
  • Operating ambient temperature: 10 to 30 °C

Compatible Accessories

  • Additional stainless steel shelves
  • Water tank with connection hose
  • USB stick with AtmoCONTROL software
  • Silicone plug for inlet port
  • Specific sample holders

Logistics

Remma and its partners provide you with serviced and configured equipment tailored to your needs. Receive your product at your facility with standard or express delivery, and an installation option, all through a single point of contact.

Warranty

Our 12/24-month warranty ensures the safety and reliability of your medical equipment throughout its lifespan.

Flexible Financing

Settle your medical equipment in up to 36 installments with our financing service or pay after delivery with Remma Facility.