• Medical Diagnostic Equipment
  • Surgical and Therapeutic Equipment
  • Medical Furniture
  • Laboratory Equipment
  • Cleaning and Sterilization
  • Spare Parts

RapidLab 1245 Blood Gas Analyser

The Siemens RapidLab 1245 is a high-performance blood gas analyser designed for high-volume clinical environments. Delivering reliable results in just 60 seconds, it accurately measures blood gases, electrolytes, metabolites, total neonatal bilirubin and provides CO-oximetry parameters. Its modular design, automatic calibration and intuitive interface make it an essential tool for laboratories requiring speed, accuracy and simplicity. Ideal for arterial, venous or capillary whole blood analysis, it contributes to rapid and effective diagnosis in intensive care or emergency settings.



Description

Comprehensive analysis – Optimised clinical performance with the Siemens RapidLab 1245

The Siemens RapidLab 1245 is a blood gas and CO-oximetry analyser designed to meet the growing needs of modern clinical laboratories. Part of the RAPIDLab 1200 series, this device incorporates state-of-the-art technology to deliver reliable and fast results while minimising manual intervention and handling errors. Its ability to process a large number of samples in a short time makes it an indispensable tool in busy departments such as emergency rooms, intensive care units and hospital laboratories.

With its user-friendly interface and automated calibration protocols, the RapidLab 1245 ensures simple and safe operation. The device's comprehensive menu allows simultaneous measurement of critical parameters such as pH, blood gases (pCO₂, pO₂), electrolytes (Na⁺, K⁺, Ca²⁺, Cl⁻), metabolites (glucose, lactate), and CO-oximetry values (O₂Hb, COHb, MetHb, HHb, tHb). It also stands out for its ability to analyse total neonatal bilirubin, an essential feature for paediatric care.

The instrument uses heparinised whole blood, collected either by syringe or capillary, giving it great clinical flexibility. It requires a relatively small sample volume (140 µL for the full menu) and provides results in 60 seconds. Its measuring chamber is designed to maximise accuracy while reducing maintenance requirements. The reagent cartridge has a shelf life of approximately 30 days, ensuring economical and consistent operation.

The RapidLab 1245 features an automatic quality control (AQC) system, ensuring optimal reliability of results. Calibration procedures are performed at regular intervals (every 30 to 60 minutes for spot calibration and every 8 hours for full calibration), ensuring a high level of compliance with regulatory requirements.

Connectivity options include USB ports, RS232, a barcode reader and an Ethernet interface, facilitating integration of the system into the laboratory's IT infrastructure. Robust and compact, the analyser measures approximately 58 x 56 x 57 cm and weighs 29.5 kg, allowing for easy installation even in confined spaces.

The RapidLab 1245 has been designed to comply with the most stringent safety and electromagnetic standards, including UL, CSA and CE certifications. It operates over a wide range of temperatures and humidity levels, ensuring reliability in a variety of environmental conditions.

The Siemens RapidLab 1245 offers an ideal combination of speed, reliability and versatility, meeting the most demanding clinical needs. Its open architecture, integrated quality management system and extensive analysis capabilities make it a strategic choice for improving patient care in critical environments.


Features

  • Rapid and comprehensive analysis of blood gases, electrolytes, metabolites and CO oximetry
  • Integration of neonatal total bilirubin measurement
  • Intuitive user interface and colour screen
  • Automated calibration and quality control
  • High processing capacity for intensive clinical environments
  • Low sample volume required
  • Compact design for easy installation
  • Easy network integration via Ethernet and multiple ports
  • Reduced maintenance thanks to long-life consumables
  • Compliance with international safety and performance standards

Technical Details

  • Parameters analysed: pH, pCO₂, pO₂, Na⁺, K⁺, Ca²⁺, Cl⁻, glucose, lactate, total neonatal bilirubin, CO oximetry (O₂Hb, HHb, COHb, MetHb, tHb, sO₂)
  • Sample type: Heparinised whole blood (syringe or capillary)
  • Result time: 60 seconds
  • Sample volume:
    • pH alone: 35 µL
    • CO oximetry alone: 100 µL
    • Full menu: 140 µL
  • Reactive cartridge shelf life: 30 days
  • Calibration:
    • 1 point every 30 to 60 minutes
    • Full calibration every 8 hours
  • Quality control: Automated AQC cartridge, QC ampoule
  • Dimensions: 58.4 × 55.9 × 57.2 cm
  • Weight: 29.5 kg
  • Power supply:
    • Voltage: 100 to 240 VAC
    • Frequency: 50–60 Hz
    • Power: 150 VA
  • Operating conditions:
    • Temperature: 15°C to 32°C
    • Humidity: 5% to 85% non-condensing
  • External interfaces: USB, RS232, barcode reader, Ethernet RJ10/100
  • Standards and safety: EN 60601-1-2, UL, CSA, FCC47-15-B


Compatible Accessories

  • Reagent cartridges 30 days
  • Quality control cartridge (AQC)
  • Barcode reader
  • External printer (USB)
  • Compatible syringes and capillaries
  • Maintenance kit
  • RJ45 Ethernet cable

Logistics

Remma and its partners provide you with serviced and configured equipment tailored to your needs. Receive your product at your facility with standard or express delivery, and an installation option, all through a single point of contact.

Warranty

Our 12/24-month warranty ensures the safety and reliability of your medical equipment throughout its lifespan.

Flexible Financing

Settle your medical equipment in up to 36 installments with our financing service or pay after delivery with Remma Facility.