Separation Module 2695

Brand: Waters

Be notified by email when the product is available
Subscribe to our partner program and become a Remma seller.


The Waters 2695 separations module is designed to integrate solvent and sample management within HPLC systems. This automated device supports up to 120 sample vials and ensures precise control of injection, flow rate, gradient formation, and temperature. Compatible with various software platforms such as Millennium and MassLynx, it enables smooth and reliable execution of chromatographic methods. With advanced features like in-line degassing, programmable needle wash, and method storage via floppy disk, the 2695 offers a comprehensive solution for demanding laboratory environments.

Description

Waters 2695: Integrated Platform for HPLC Chromatography

The Waters 2695 is an all-in-one system that combines solvent and sample management to automate high-performance liquid chromatography (HPLC) analyses. By centralizing critical separation functions within an integrated platform, it minimizes manual handling while maximizing result reliability.

This module controls up to four solvents and delivers exceptional accuracy in gradient management, flow rate, and injection volume. It supports method programming, solvent degassing, column and sample temperature control (with optional modules), and can interface via RS-232 or IEEE-488, depending on the chosen configuration.

The 2695 accommodates up to 120 vials distributed across five carousels, with a configurable injection range of 1 to 99 per sample. Its user interface—featuring an LCD screen and keypad—makes it easy to configure methods and monitor operations in real time. A built-in floppy disk drive facilitates method transfer and archiving, enabling streamlined deployment of protocols across multiple systems.

The module also ensures complete traceability by automatically logging key operational data such as total number of injections, solvent consumption, and system errors. Optional accessories include a barcode reader, sample heating/cooling module, and PerformancePLUS™ in-line degasser to meet specific analytical requirements.

In terms of compatibility, the Waters 2695 can operate as a standalone unit or be integrated into complex systems, communicating with a wide range of chromatographic software and instrumentation. It also supports multiple injection strategies, including standard and extended loop configurations.

Reliable, precise, and modular, the Waters 2695 is a robust solution for research, quality control, or analytical development laboratories seeking a high-performance chromatography system adaptable to diverse applications.

Features

  • Full integration of solvent and sample management
  • Seamless interface with Millennium® and MassLynx™ software
  • LCD display and keypad for intuitive local control
  • In-line degassing for enhanced solvent stability and baseline performance
  • Floppy disk storage and transfer for easy method portability
  • Automated data logging: injections, solvent usage, system errors
  • Programmable needle and seal wash for improved injection reproducibility
  • Flexible injection modes with precise volume control
  • RS-232, IEEE-488, and I/O ports for easy system integration
  • Functionalities: stat runs, automatic dilutions, internal standards support

Technical Details

  • Dimensions: H 57.1 cm × W 45.7 cm × D 57.1 cm (up to 64.8 cm with optional modules)
  • Weight: 45.5 kg
  • Operating Temperature: 4°C to 40°C
  • Relative Humidity: 20% to 80% (non-condensing)
  • Maximum Operating Pressure: 5000 psi (345 bar)
  • Flow Rate Accuracy: ≤ 0.075% RSD
  • Programmable Flow Rate: 0.000 to 10.000 mL/min (increments of 0.001 mL/min)
  • Composition Range: 0–100% (in 0.1% increments)
  • System Dwell Volume: < 650 µL
  • Injection Volume: 0.1 to 100 µL (up to 2000 µL with optional loop)
  • Sample Capacity: 120 vials (5 carousels of 24)
  • Power Supply: 85–132 V or 180–250 V, 950 VA max, 47–63 Hz

Compatible Accessories

  • Column heater for consistent and controlled temperature conditions
  • Sample heater/chiller for precise temperature regulation of vials
  • Barcode reader for automated sample identification and tracking
  • PerformancePLUS™ degasser for enhanced solvent degassing performance
  • Column selection valves to enable automated switching between multiple columns
  • Additional sample loops to extend injection volume range or support specialized workflows

Discover more in this category

Logistics

Remma and its partners provide you with serviced and configured equipment tailored to your needs. Receive your product at your facility with standard or express delivery, and an installation option, all through a single point of contact.

Warranty

Our 12/24-month warranty ensures the safety and reliability of your medical equipment throughout its lifespan.

Flexible Financing

Settle your medical equipment in up to 36 installments with our financing service or pay after delivery with Remma Facility.