- Dimensions: 7-1/2″ W x 6-3/4″ D x 10″ H
- Weight: 10 lbs
- Systolic Pressure: Number of Digits: 3; Accuracy: Mean error less than ±5 mmHg, Standard deviation less than ±8 mmHg; Range: Adult Mode: 55 to 260 mmHg; Pediatric Mode: 55 to 160 mmHg
- Diastolic Pressure: Number of Digits: 3; Accuracy: Mean error less than ±5 mmHg, Standard deviation less than ±8 mmHg; Range: Adult Mode: 30 to 200 mmHg; Pediatric Mode: 30 to 150 mmHg
- NIBP Measurement Cycle: Less than 30 seconds average at 80 BPM with 180 mmHg pump up pressure, without retries, motion artifact, or arrhythmia with standard adult cuff on a healthy individual. Cycle time is affected by arm size and wrapping technique.
- Pulse Rate: Range: 35 – 245 BPM for Adult and Pediatric; Display Resolution: 1 BPM; Accuracy: ±3 BPM or ±3%, whichever is greater
- Maximum Cuff Pressure: Two means of limiting cuff pressure are provided; a hardware overpressure monitor which limits the pressure to 330 mmHg for adults and 220 mmHg for pediatrics. A software overpressure monitor which vents if the pressure exceeds 300 mmHg for adults and 200 mmHg for pediatrics. If the hardware overpressure circuit is tripped in normal operation, the unit must be turned off and back on to reset the system.
- Inflation Source: This inflation source is capable of supplying sufficient air to bring a volume of 700cc to a pressure of 300 mmHg in no more than 35 seconds. If the cuff is not inflated to the desired pressure within 60 seconds, the cuff is vented and a retry cycle is initiated.
- Cuff Vent Rate: When the unit is vented, a volume of at least 700cc is reduced from a pressure of 250 mmHg to a pressure of 20 mmHg in a maximum of 14 seconds.
- SpO2: Range: 40 – 100%; Display Resolution: 1%; Display Update: Every 2 seconds; Calibration: Factory calibrated to functional saturation; Accuracy: from 70 – 100% SpO2; from 60 – 69% SpO2; Unspecified from 40 – 59% SpO2
- Battery: Type: Sealed lead-acid; Quantity: 1; Voltage: 12 V DC nominal; Capacity: 2.3 Amp-Hour
- Run Time: Accutorr Plus NIBP – 5 hours from full charge with new battery at 25°C with 1 NIBP measurement every 5 minutes and recorder not in use; Accutorr Plus NIBP with Trend Screen – 3 hours from full charge with new battery at 25°C with 1 NIBP measurement every 5 minutes and recorder not in use; Accutorr Plus NIBP with Trend Screen and SpO2 – 2 hours from full charge with new battery at 25°C with 1 NIBP measurement every 5 minutes, continuous SpO2 measurement, and recorder not in use
- Recharge Time: 8 hours maximum with unit in Standby
- Cyclic Service Life: 150 cycles, 100% to 20% of capacity; 400 cycles, 100% to 40% of capacity
- Real Time Clock: Resolution: 1 minute; Accuracy: ±1 minute/week; Display Format: 24 hours; Power: The real-time clock maintains the time and date when the instrument is On or in the standby mode, connected to AC mains or running from internal battery for at least ten years from original assembly. The real-time clock will maintain time and date even if the instrument’s main battery is disconnected.