• Non-invasive blood pressure measurement using GE's exceptional DINAMAP* technology
• Initial inflation pressures can be set, allowing you to adapt to each patient's specific needs and ensure their comfort
• Large screen for easy reading, even from a distance
• Capable of storing up to 40 measurements over a period of up to 24 hours, with the option to print tables
• Designed for easy maintenance with a removable panel for easy access and easy-to-use on-site replacement kits
• NIBP (non-invasive blood pressure measurement) specifications
• Cuff pressure range 0 to 290 mmHg (adult/paediatric) (normal operating range) 0 to 145 mmHg (neonatal)
Blood pressure accuracy
• NIBP algorithm Average error ≤5 mmHg,
• Conventional and auscultatory average error ≤5 mmHg,
• Standard deviation ≤8 mmHg (compliant with ANSI/AAMI SP10:2002 standard) maximum time 120 s (adult/paediatric) determination 85 s (newborn)
• Overpressure threshold 300 to 330 mmHg (adult/paediatric) 150 to 165 mmHg (newborn)
Blood pressure range
• PNI SuperSTAT algorithm Systolic 30 to 290 mmHg (adult/paediatric) 30 to 140 mmHg (neonatal)
• MAP 20 to 260 mmHg (adult/paediatric) 20 to 125 mmHg (neonatal)
• Diastolic 10 to 220 mmHg (adult/paediatric) 10 to 110 mmHg (newborn)
• Classic and auscultatory Systolic 30 to 245 mmHg (adult/paediatric) 40 to 140 mmHg (newborn) MAP 15 to 215 mmHg (adult/paediatric) 30 to 115 mmHg (newborn)
• Diastolic 10 to 195 mmHg (adult/paediatric) 20 to 100 mmHg (newborn)
Accessories :
With mobile trolley and temperature probe