- Dimensions: 26 cm W x 22 cm H x 14.5 cm D (10.2" W x 8.6" H x 5.7" D)
- Weight: 3.0 kg (6.6 lbs) excluding optional recorder and temperature module
- Screen type: 21.3 cm (8.4") SVGA TFT-AM LCD display
- Display resolution: 800 active pixels/line, 600 active lines/frame
- Screen active area: 170.4 mm x 127.8 mm (6.71" x 5.03")
- Screen refresh frequency: 60 Hz
- Pixel size: 0.213 mm (inch)
- Viewing angle: approx. 60 °
- Alarm Audio Range: 45 - 85 dB
- System Response Time: 1 s
- Recorder: Recorder type: Thermal
- Paper width: 58 mm (inch)
- User Selectable Speeds: 6.25, 12.5, 25, and 50 mm/s
- Electrical specifications: Internal battery: Lithium ion, Smart battery 10.9 V to 11.1 V, 7200 mAhr
- Battery operating time (new, fully charged): 4 hour(s)
- Battery charge time: <4 hour(s)
- Internal power supply: 100 to 240 VAC
- Power consumption: 72 W
- Frequency: 50/60 Hz
- Environmental specifications: Mechanical Shock: Complies with mechanical shock requirement in ISO 9919 standards for use within the healthcare facility.
- Mechanical Vibration: Complies with mechanical vibration requirement in ISO 9919 standards for use within the healthcare facility.
- Operating temperature: 10 to 40 °C (50 to 104 °F)
- Storage temperature (device): -20 to 50°C (-4 to 122 °F)
- Storage temperature (device plus accessories): -20 to 40°C (-4 to 104 °F)
- Operating humidity: Up to 80% RH, non-condensing
- Storage humidity: Up to 90% RH, non-condensing
- Altitude: Up to 3000 meters above sea level
- ECG: Heart rate range: 15 to 300 beats per minute
- Heart rate accuracy: Greater of ±10% or ±5 bpm
- Bandwidth (normal monitoring): 0.67 to 40 Hz
- Bandwidth (filtered monitoring): 0.67 to 20 Hz
- Leads: 3-lead
- Display sweep speeds: 12.5, 25, and 50 mm/s
- Pacemaker detection: Indicator on waveform display (user-selectable)
- ECG size (sensitivity): 2.0, 1.0, 0.5 or 0.25 cm/mV or Auto
- Lead Off condition: Detected and displayed
- Impedance Respiration: Technique: Trans-thoracic impedance
- Measurement range: 0 to 150 rpm
- Accuracy: ±1 rpm in the range of 0 to 120 rpm; ±2 rpm in the range of >120 rpm
- ECG leads used: RA to LL
- Display sweep speeds: 6.25, 12.5, 25, or 50 mm/s
- Lead Off condition: Detected and displayed
- Non-invasive Blood Pressure: Technique: Oscillometric using stepwise deflation pressure
- Adult measurement range - systolic: 30 mmHg to 270 mmHg (4.0 kPa to 36.0 kPa)
- Initial cuff inflation - Adult: 160 mmHg (21.3 kPa)
- Adult measurement range - diastolic: 10 mmHg to 245 mmHg (1.3 kPa to 32.7 kPa)
- Adult measurement range - MAP: 20 mmHg to 255 mmHg (2.7 kPa to 34.0 kPa)
- Adult measurement range - pulse rate: 40 to 300 beats per minute
- Pediatric measurement range - systolic: 30 mmHg to 180 mmHg (4.0 kPa to 24.0 kPa)
- Initial cuff inflation - Pediatric: 140 mmHg (18.7 kPa)
- Pediatric measurement range - diastolic: 10 mmHg to 150 mmHg (1.3 kPa to 20.0 kPa)
- Pediatric measurement range - MAP: 20 mmHg to 160 mmHg (2.7 kPa to 21.3 kPa)
- Pediatric measurement range - pulse rate: 40 to 300 beats per minute
- Neonatal measurement range - systolic: 30 mmHg to 130 mmHg (4.0 kPa to 17.0 kPa)
- Neonatal measurement range - diastolic: 10 mmHg to 100 mmHg (1.3 kPa to 13.3 kPa)
- Neonatal measurement range - MAP: 20 mmHg to 120 mmHg (2.7 kPa to 16.0 kPa)
- Neonatal measurement range - pulse rate: 40 to 300 beats per minute
- Blood pressure accuracy: 8 mmHg maximum standard deviation; ±5 mmHg maximum mean error
- Pulse rate accuracy: 40 to 100 bpm: ±5 bpm; 101 to 200 bpm: ±5% of reading; 201 to 300 bpm: ±10% of reading
- Initial cuff inflation - neonatal: 100 mmHg (13.3 kPa)
- NBP intervals: Automatic measurements at intervals of 1, 2, 3, 5, 10, 15, 30, 60, 90, 120 minutes, and STAT
- Temperature Measurements: Measurement range - monitored mode: 26.7 to 43.3 °C (80 to 110 °F)
- Measurement range - predictive mode: 34.4 to 40.6 °C (93.9 to 105 °F)
- Accuracy (monitored mode): ±0.1 °C (±0.2 °F)
- Heating/Cooling transient response time: ≤150 s
- Measurement time: 4 s (Oral); <16 s (Axillary and rectal) s
- Resolution: 0.1 °C (0.2 °F)
- SpO2: Measurement range: 0 to 100 %
- Pulse rate: 30 to 300 beats per minute
- SpO2 accuracy: 70 to 100% (exact accuracy depends on sensor) %
- Pulse rate accuracy: Greater of 2% or ±1 bpm
- Wavelength range: 500 to 1000 nm for all specified sensors nm
- Maximum optical output power: ≤15 mW for all specified sensors mW